The MBI Nano-Spectrophotometer 300 is a full-spectrum, UV-Vis micro-spectrophotometer covering 200–800 nm. It requires only 0.5–2 µL of sample to rapidly and accurately detect nucleic acids, proteins, and cell solutions — with results displayed in under 5 seconds.
Equipped with a 7-inch capacitive touch screen running on a customised Android system, the Nano-300 needs no external computer. It features a cuvette mode for measuring bacteria and other culture media concentrations, plus an OD600 function for bacterial cell density estimation. Data can be printed via the built-in printer or exported via USB for analysis and storage.
Key Features
• Android system with 7-inch capacitive touch screen
• Full spectrum 200–800 nm scan in under 5 seconds
• Micro-volume mode: only 0.5–2 µL sample required
• Cuvette mode for bacteria and culture media concentration
• OD600 function for bacterial cell density estimation
• Built-in printer for instant hard-copy reports
• USB data export for analysis and storage
• No lamp warm-up — instant measurement upon power-on
Application Range
• 260 nm — dsDNA, ssDNA, RNA quantification
• 280 nm — A280, BSA, IgG, Lysozyme
• 595 nm — Bradford protein assay
• 562 nm — BCA protein assay
• 650 nm — Lowry protein assay
• 600 nm — Bacterial cell concentration (OD600)
How to Use
1. Power on — No warm-up required; the device is ready instantly.
2. Load sample — Pipette 0.5–2 µL onto the micro-volume pedestal, or insert a cuvette for OD600/culture media.
3. Measure — Lower the detection arm; the device auto-detects and displays results in under 5 seconds.
4. Output — Print via built-in printer or export data via USB flash drive.
Technical Specifications
Parameter
Value
SKU
MBIZR-00919-99
Product Name
MBI Nano-Spectrophotometer 300
Wavelength Range
200–800 nm
Minimum Sample Size
0.5–2.0 µL
Path Length
0.2 mm / 1.0 mm (auto-ranging)
Light Source
Xenon flash lamp
Detector Type
2048-element linear CCD array
Wavelength Accuracy
±1 nm
Spectral Resolution
≤3 nm (FWHM at Hg 546 nm)
Absorbance Precision
0.003 Abs
Absorbance Accuracy
1% (7.332 Abs at 260 nm)
Absorbance Range
0.04–90 A
Nucleic Acid Detection Range
2–4500 ng/µL (dsDNA)
Measurement Time
<5 seconds
Data Output
Built-in printer + USB flash drive
Sample Pedestal Material
Aluminum alloy and quartz fiber
Operating Voltage
DC 24 V, 2 A
Operating Power
25 W
Standby Power
5 W
Dimensions (W × D × H)
210 × 268 × 181 mm
Weight
2.8 kg
Software Compatibility
Android system (built-in)
Auto Sleep
Screen turns off after 5 minutes idle; wake by touch