MBI Evolution Microplate Reader (340-750nm)
|
```
| Light Source | Quartz-halogen lamp |
|---|---|
| Measurement range | 0.000…4.000 OD |
| Precision | CV≤0.2% (0-3Abs) CV≤1.0% (3-4Abs) at 405nm |
| Accuracy | ±1% (0-3Abs) ±2% (3-4Abs) at 405nm |
| Standard filters | 405, 450, 492, 630 nm |
| Wavelength Selection | Filters |
| Half-Bandwidth of Filters | 3~9nm |
| Wavelength range | 340 - 750nm |
| Test Speed | 6 seconds for 96 wells plate |
| Detector | Highly sensitive channel photo multiplier (CPM) - optimal signal/noise ration by effective noise reduction (decreased SNR (signal/noise ratio)-technology) |
| Shaking | 3 modes shaking: slow, medium and fast |
| Optical system | Filters |
| Linearity (±) | <±0.25% and ±0.0025 OD from 0.1 OD to 2.5 OD at 492 nm |
| Data Output | 3 USB ports, for PC, printer and USB-disk |
| Dimensions | 440 x 295 x 225 mm |
| Net Weight | 10.0 Kg |
| PC Operation | Yes |
| Microplate Formats | 96 well plate (flat, round and v-bottomed well formats) |