Features

  • Split Beam (RBC) UV/Vis Spectrophotometer covering 190-1100nm (3/5nm bandwidth)
  • Compact, and excellent value for money compared to conventional systems.
  • Wavelength scanning, kinetics and concentration functionality with full graphics display
  • Flexible data export optional accessories include Export to PC (PVC) software,  integrated printer*, wireless Bluetooth® connectivity*, SD card accessory*, USB cable*

 

Compact and low maintainance: The Biochrom WPA Lightwave combines ease of use with flexibility, and the novel optics require no lid or moving parts  and use a Xenon source for high energy performance with longer lamp lifetime. Compact footprint saves valuable bench space. This instrument has a 2 year warranty and the Xenon lamp has a 3 year warranty. The Lightwave II+ is a higher resolution 3nm bandwidth version. 

 

Powerful and flexible: The instrument includes a large wide view display and in-built software providing flash scan, fixed wavelength measurements, kinetics and concentration with comprehensive graphics capability plus the ability to store up to 90 methods. On peak confirmation is also a feature of the flexible software. Concentration may be measured using either a factor, singlepoint calibration, multi-standard curves or alternatively there is a multi-wavelength mode where equations using absorbance values may be used for ratio calculations. Samples may be measured in 10, 20 or 40mm pathlength cells (glass, quartz or disposable).

 

Flexible data export options: Results may be printed to an optional integrated high quality graphical printer for a permanent record or exported via a USB connection to a suitable PC running Print Via Computer (PVC) software for advanced reporting and data storage or the optional wireless Bluetooth® accessory. Alternatively the optional SD card accessory allows laboratories to export to standard SD or SDHC memory cards. This will be particularly useful to laboratories wishing to clone methods and folders between instruments. Please note that only one optional accessory can be fitted to the instrument.

 

Upgrade and Spare parts

  • 80-3004-07 Spare printer paper (20 rolls).
  • 80-3003-84 Printer accessory.
  • 80-3004-73 Print via computer software and cable.
  • 80-3003-96 Bluetooth accessory
  • 80-3005-00 SD Card accessory.

Full Technical specifications are shown in additional information. Optional accessories marked *

More Information
Spectrophotometer type UV / Visible
Wavelength range 190 - 1100nm
Wavelength Accuracy ± 2nm
Wavelength Reproducibility ± 1nm
Light Source Xenon (PTR)
Beam Split Beam
Optical system Monochromator Holographic Grating
Bandwidth 5nm (3nm Lightwave II+)
Detector 1024 CCD array
Photometric Range -0.300 -> 2.500A; 0 -> 199%T
Photometric accuracy +/- 0.008Abs or 1.3% of the reading, whichever is greater between 0 and 2.5A
Photometric reproducibility ± 0.002A at 0-0.5A, 546nm
Stray Light 0.5%T at 220 and 340nm
Stability 0.001A/h at 340nm
Measurement Type Fixed wavelength (Abs/%T), Abs ratio, concentration (standard curve), concentration (factor), multiwavelength, kinetics
Cell Compatibility Cuvette, Microcell*, Long Pathlength Cell*
Cuvette holder Single Cell
Accessories Printer*
Methods Up to 90
Regulatory Compliance CE
Data Output Print*, Store*, Export to PC*
Display Graphics LCD
Data interface USB
Software Languages (on-board software) English
Power requirements 90-250 V AC, 50/60Hz, 30VA
Net Weight 4.5Kg (9.9lb)
Dimensions 10 x 26 x 39 cm (3.9 x 10.2 x 15.3 inches)