My Cart
My Quotes

Spectrophotometer : Lightwave 3 with Printer, 5nm bandwith

Catalog Number:
80-3007-43
Availability: Out of stock
Please contact us

Lightwave 3 with PrinterUV-Vis Spectrophotometer, 5nm bandwidth, USB port for data collection, in-built serial printer.

Features

  • Split Beam (RBC) UV/Vis Spectrophotometer covering 190-1100nm (3/5nm bandwidth)
  • Accepts 10, 20 and 40mm pathlength cuvettes
  • Compact, and excellent value for money compared to conventional systems.
  • Wavelength scanning, kinetics and concentration functionality with full graphics display

Compact and low maintenance
The Biochrom WPA Lightwave combines ease of use with flexibility, and the novel optics require no lid or moving parts and use a Xenon source for high energy performance with longer lamp lifetime. Compact footprint saves valuable bench space. This instrument has a 2 year warranty and the Xenon lamp has a 3 year warranty. The Lightwave 3+ is a higher resolution 3nm bandwidth version. 

Powerful and flexible
The instrument includes a large wide view display and in-built software providing flash scan, fixed wavelength measurements, kinetics and concentration with comprehensive graphics capability plus the ability to store up to 90 methods. On peak confirmation is also a feature of the flexible software. Concentration may be measured using either a factor, single point calibration, multi-standard curves or alternatively there is a multi-wavelength mode where equations using absorbance values may be used for ratio calculations. Samples may be measured in 10, 20 or 40mm pathlength cells (glass, quartz or disposable).

Flexible data export options
All instruments have a USB port to download result data to a flash drive. Alternatively, results may be printed to an optional integrated graphical printer for a permanent record or exported via a USB cable connection to a suitable PC running Print Via Computer (PVC) software for advanced reporting and data storage or the optional wireless Bluetooth® accessory.

SpecificationData
Wavelength Range 190 to 1100 nm
Monochromator Flat grating
Wavelength Calibration Automatic upon switch on
Beam Height 15 mm
Spectral Bandwidth 5 nm or 3 nm for ‘+’ models
Wavelength Accuracy ±2 nm
Wavelength Reproducibility ±1 nm
Light Source Xenon flash lamp
Detector Twin CMOS array
Photometric Range -0.300 to 2.500 A, 0.3 to 199 %T
Photometric Linearity ±1.3 % or ±0.008 A whichever is greater at 546 nm
Photometric Reproducibility ±0.002 A to 0.5 A at 546 nm
Stray Light <0.5 %T 340 nm
Stability ±0.01 A/h at 340 nm
Noise ±0.005 peak to peak ±0.002 RMS
Dimensions 260 × 390 × 100 mm
Weight 3.00 kg 3.00 kg


 
More Information
Wavelength range 190 - 1100nm
Wavelength Accuracy ± 2nm
Wavelength Reproducibility ± 1nm
Light Source Xenon flash lamp
Stray Light <0.05%T at 340nm
Temperature Control -
Net Weight 3.0 kg
Write Your Own Review
You're reviewing:Spectrophotometer : Lightwave 3 with Printer, 5nm bandwith