Lab Companion (Jeio Tech)

Vertical type laminar flow clean bench

  • ISO 14644-1 Class 4 air (formerly Class 10 per Federal Standard 209E)
  • Supply HEPA filter, 99.99% efficient on particles 0.3micron
  • Nominal downflow velocity of 0.3m/s (60fpm)
  • Ergonomic 5 degree sloped front for comfortable use

 

Features

Particle-free ISO 14644-1 Class 4 air

  • Particle count in the work area conform to ISO Class 4 conditions (formerly Class 10 per Federal Standard 209E). This ensures fewer than 0.3um or larger per cubicmeter or air
  • Polyester fiber pre-filter that has a small pressure loss and 85% arrestance on the A.F.L. test which serves to trap larger particles and increase the life of the main filter
  • HEPA filter ensures airtight to the body unit by connecting with neoprene gaskets * The HEPA filter life span is usually 3 years, but it depends on test room conditions

 

Ergonomic design

  • Angled front profile is free of protrusions that could interfere with visibility or obstruct taller users
  • 304 stainless steel work surface - Highly durable and rust prevented work zone for easy cleaning
  • Use of tempered glass door to be strong of damage and intercept the UV light for user safety
  • The door smoothly slides up and down for maintaining cleanness and stops at any height you want
  • Right and left both sides also have wide glasses for a clear view
  • Adjustable the speed of blower by front located blower controller

 

Lighting

  • Electronically ballasted FL lamp is installed on the top center that minimizes disturbance of airflow and shadows on the work zone, and to relieve of user's eyes strain

 

Clean Bench Airflow Profile

  • Room air is taken in from the top blower of the bench through a disposable pre-filter
  • The firstly filtered air is purified by the secondary HEPA filter to confirm the class
  • The laminar down-flowed air supplies from the dual filtering system into the working zone
  • The working zone can maintain on a high cleanness condition because it keeps up on positive pressure to prevent from outside pollution
  • The purified air is across to the working zone of the cabinet in a vertical, unidirectional stream and leaves the main work chamber across the entire open front of the cabinet and through slots

Note: Clean Bench is not a Class II, Type A2 Biological Safety Cabinet. As such, it should not be used for cell cultures or when products manipulated could be nocive for the user.

 

More Information
Airflow Type Vertical
Air Volume 274 to 914 cmh/ 119 to 399 cfm
Laminar Air Velocity 0.3 / 0.6
Air Cleanliness within Working Area ISO 14644-1 Class 4, US Federal Standard 209E Class 10
HEPA Filter Typical efficiency of 99.99% at 0.3 um US MIL-STD-282 ; Micro glass fiber Media, Particle board, AL separator, Neoprene gaskets
Pre-Filter Polyester fibers with an efficiency of 85% (A∙F∙I TEST) ; AL frame, Polyester fiber Media
Noise Level Typically <67 dBA at initial Blower speed
Body material Stainless steel grade 304 & Steel powder coating
Working Surface Materials Stainless steel grade 304
Front/ Side Windows Materials Colourless and transparent UV absorbing 5mm tempered glass
Illumination Intensity >1000
Illumination FL Lamp 30W×2EA electronically ballasted FL Lamp
Illumination UV Lamp 30W electronically ballasted UV Lamp
Electric socket Outlets 230V socket, 120V socket
Exterior Dimensions 1040×630×1060 / 40.1×24.8×42.7
Interior Dimensions 970×540×570 / 38.2×21.3×23.2
Net Weight 110 / 242.5 (kg/lbs)
Power Supply 60Hz, 1.33A, Iph (160W)