My Cart
My Quotes

Flame Photometer for the Laboratory FP8400

Catalog Number:
FP8400
Availability: In stock
Please contact us

Flame Photometer for the Laboratory FP8400

Economic entry-level model for the laboratory

The flame photometer FP8400 is the standard version for a variety of laboratory applications and offers a very simple operation in combination with very high precision and application flexibility in determining the concentration of alkali and alkaline earth elements in aqueous solutions. 

With the models of the FP8000 series, A.KRÜSS Optronic presents to this day unique measuring devices for the simultaneous measurement of the elements Na, K, Ca and Li. Five measuring channels are available for this, which can be individually configured. Thanks to proven measuring methods and modern device electronics, our flame photometers achieve high-precision and reproducible measurement results and are a simple and particularly cost-effective alternative to analysis techniques such as ICP or AAS.

  • Reliable, simultaneous measurement of up to five alkaline and alkaline earth elements
  • For operation with low sample volume
  • Very high precision
  • Maximum operational safety through intelligent security mechanisms
  • User administration with two authorization levels
  • Full traceability of the measurement results
  • Extensive interfaces and convenient data transfer
  • Compliance with international standards such as GMP / GLP and 21 CFR Part 11

 

Typical applications

  • Quality control in the pharmaceutical, chemical, beverage, food, salt and construction industries
  • Evidence
  • Food laboratories, testing laboratories and testing offices
  • Chemical development
 

 

Model Specifications

FP8400

Na

K

Li

Ca

MEASURING RANGE

0.01-4500 ppm


0.0004-200 mmol/l

0.01-4500 ppm


0.0003-110 mmol/l

0.01-4500 ppm


0.0014-600mmol/l

0.50-4500 ppm


0.0125-110 mmol/l

DETECTION LIMIT

0.01 ppm


0.0004 mmol/l

0.01 ppm


0.0003 mmol/l

0.01 ppm


0.0014 mmol/l

0.03 ppm


0.0075 mmol/l

PRECISION

0.2 % at 40 ppm


0.2 % at 1,74 mmol/l

0.2 % at 40 ppm


0.2 % at 1,03 mmol/l

0.2 % at 40 ppm


0.2 % at 5.71 mmol/l

0.2 % at 40 ppm


0.2 % at 1.00 mmol/l

ACCURACY

1 % at 40 ppm


1 % at 1.74 mmol/l

1 % bei 40 ppm


1 % at 1.03 mmol/l

1 % at 40 ppm


1 % at 5.71 mmol/l

1 % at 40 ppm


1 % at 1.00 mmol/l

CALIBRATION

  • Linear with 2 standards
  • Nonlinear with 6-8 standards,
  • Cubic approximation

DRIFT

1% in 60 min

REFERENCE

Lithium guide line with 35 mg/l or 5 mmol/l

SAMPLE VOLUME

2.5 ml

COMBUSTION GAS

Propane, Butane or LPG: Recommended for alkaline elements


Acetylene for flame photometry: Recommended for alkaline earth elements

DISPLAY

TFT-Display with integrated 8,4“TFT touchscreen, 800x600 Pixel

INTERFACES

2 x USB
1 x Ethernet
1 x RS-232 for printer

EQUIPMENT

Basic model

ACCESSORIES

  • 24-character plain paper printer
  • Pressure reducing station with filter unit
  • Various standards and system solutions
  • Spare parts and accessories directly from the manufacturer

UPGRADEABILITY

Retrofit kits to other models available

 

More Information
MEASURING RANGE Na 0.01-4500 ppm
MEASURING RANGE K 0.01-4500 ppm
MEASURING RANGE Li 0.01-4500 ppm
MEASURING RANGE Ca 0.50-4500 ppm
DETECTION LIMIT Na 0.01 ppm
DETECTION LIMIT K 0.01 ppm
DETECTION LIMIT Li 0.01 ppm
DETECTION LIMIT Ca 0.03 ppm
PRECISION Na 0.2 % at 40 ppm
PRECISION K 0.2 % at 40 ppm
PRECISION Li 0.2 % at 40 ppm
PRECISION Ca 0.2 % at 40 ppm
ACCURACY Na 1 % at 40 ppm
ACCURACY K 1 % at 40 ppm
ACCURACY Li 1 % at 40 ppm
ACCURACY Ca 1 % at 40 ppm
Calibration Linear with 2 standards Non linear with 6-8 standards, cubic approximation
Sample volume 2,5 ml
Display TFT-Display with integrated 8,4“ TFT touchscreen, 800x600 Pixel
Interface 2 x USB flash drive 1 x Ethernet 1 x RS-232 for printer
Write Your Own Review
You're reviewing:Flame Photometer for the Laboratory FP8400
flamephoto (1.11 MB)