Flame Photometer - Automatic Unit With Dilution FP8700
Flame Photometer - Automatic Unit with Dilution FP8700
Automated process for operation with high sample throughput
The flame photometer FP8700 is the laboratory version supplemented by a rotation sampler, diluter, and mixer. Ideal for applications with high sample volumes and high concentrations requiring dilution. Various safety mechanisms ensure particularly high operational safety. Especially control laboratories and testing offices benefit from this independent operation.
With the models of the FP8000 series, A.KRÜSS Optronic presents to this day unique measuring devices for the simultaneous measurement of the elements Na, K, Ca and Li. Five measuring channels are available for this, which can be individually configured. Thanks to proven measuring methods and modern device electronics, our flame photometers achieve high-precision and reproducible measurement results and are a simple and particularly cost-effective alternative to analysis techniques such as ICP or AAS.
- Reliable, simultaneous measurement of up to five alkaline and alkaline earth elements
- Automatic dilution
- Automatic sample feed
- Suitable for high sample throughput and small sample volumes
- Little care needed by users
- Very high precision
- Maximum operational safety through intelligent security mechanisms
- User administration with two authorization levels
- Full traceability of the measurement results
- Extensive interfaces and convenient data transfer
- Compliance with international standards such as GMP / GLP and 21 CFR Part 11
Typical applications
- Quality control in the pharmaceutical, chemical, beverage, food, salt and construction industries
- Evidence
- Food laboratories, testing laboratories and testing offices
- Chemical development
Model Specifications
|
FP8700 |
Na |
K |
Li |
Ca |
|
MEASURING RANGE |
0.1-45000 ppm
|
0.1-45000 ppm
|
0.1-45000 ppm
|
5.0-45000 ppm
|
|
DETECTION LIMIT |
0.1 ppm
|
0.1 ppm
|
0.1 ppm
|
0.3 ppm
|
|
PRECISION |
0.6 % at 40 ppm
|
0.6 % at 40 ppm
|
0.6 % at 40 ppm
|
0.6 % at 40 ppm
|
|
ACCURACY |
1 % at 40 ppm
|
1 % at 40 ppm
|
1 % at 40 ppm
|
1 % at 40 ppm
|
|
CALIBRATION |
UPGRADEABILITY |
|||
|
DRIFT |
1% in 60 min |
|||
|
REFERENCE |
Lithium guide line with 35 mg/l or 5 mmol/l |
|||
|
SAMPLE VOLUME |
0.25 ml |
|||
|
COMBUSTION GAS |
Propane, Butane or LPG: Recommended for alkaline elements
|
|||
|
DISPLAY |
TFT-Display with integrated 8,4“ TFT touchscreen, 800x600 Pixel |
|||
|
INTERFACES |
|
|||
|
EQUIPMENT |
Basic unit + External automation system, consisting of: Sampler, Diluter and Mixer |
|||
|
ACCESSORIES |
External automation system, consisting of: Sampler, Diluter and Mixer |
|||
|
UPGRADEABILITY |
Retrofit kits to other models available |
|||
| MEASURING RANGE Na | 0.1-45000 ppm |
|---|---|
| Plage de mesure K | 0.1-45000 ppm |
| MEASURING RANGE Li | 0.1-45000 ppm |
| MEASURING RANGE Ca | 5.0-45000 ppm |
| LIMITES DE DÉTECTION Na | 0.1 ppm |
| LIMITES DE DÉTECTION K | 0.1 ppm |
| LIMITES DE DÉTECTION Li | 0.1 ppm |
| LIMITES DE DÉTECTION Ca | 0.3 ppm |
| Précision du sodium. | 0.6 % at 40 ppm |
| Précision du potassium. | 0.6 % at 40 ppm |
| Précision du lithium. | 0.6 % at 40 ppm |
| Précision du calcium. | 0,6 % à 40 ppm. |
| PRÉCISION Na | 1 % à 40 ppm |
| PRÉCISION K | 1 % à 40 ppm |
| PRÉCISION Li | 1 % à 40 ppm |
| PRÉCISION Ca | 1 % à 40 ppm |
| Calibration | Linéaire avec 2 étalons. Non linéaire avec 6-8 étalons, approximation cubique. |
| Volume d'échantillon | 2,5 ml |
| Écran | "Écran TFT intégré de 8,4 pouces avec écran tactile, 800x600 pixels |
| Interface | 2 x clé USB 1 x Ethernet 1 x RS-232 pour imprimante |