Mon panier
Mes devis

Flame Photometer for Process FP8500

Numéro de catalogue:
FP8500
Disponibilité: En stock
Please contact us

Flame Photometer for Process FP8500

Process version

The FP8500 combines the characteristics of the FP8400 with an automatic calibration to realize a 24 hour fully automated measurement operation; measuring times between 1 second and several minutes are possible. For data transmission, a variety of interfaces are available. 

With the models of the FP8000 series, A.KRÜSS Optronic presents to this day unique measuring devices for the simultaneous measurement of the elements Na, K, Ca and Li. Five measuring channels are available for this, which can be individually configured. Thanks to proven measuring methods and modern device electronics, our flame photometers achieve high-precision and reproducible measurement results and are a simple and particularly cost-effective alternative to analysis techniques such as ICP or AAS.

  • Reliable, simultaneous measurement of up to five alkaline and alkaline earth elements
  • Remote control of several devices via central process computer
  • Suitable for 24-hour operation
  • Very high precision
  • Maximum operational safety through intelligent security mechanisms
  • User administration with two authorization levels
  • Full traceability of the measurement results
  • Extensive interfaces and convenient data transfer
  • Compliance with international standards such as GMP / GLP and 21 CFR Part 11

 

Typical applications

  • Process analysis
  • Process monitoring
  • Online measurement
  • Automated continuous measurement

FP8500

Na

K

Li

Ca

MEASURING RANGE

0.01-4500 ppm


0.0004-200 mmol/l

0.01-4500 ppm


0.0003-110 mmol/l

0.01-4500 ppm


0.0014-600mmol/l

0.50-4500 ppm


0.0125-110 mmol/l

DETECTION LIMIT

0.01 ppm


0.0004 mmol/l

0.01 ppm


0.0003 mmol/l

0.01 ppm


0.0014 mmol/l

0.03 ppm


0.0075 mmol/l

PRECISION

0.2 % at 40 ppm


0.2 % at 1,74 mmol/l

0.2 % at 40 ppm


0.2 % at 1,03 mmol/l

0.2 % at 40 ppm


0.2 % at 5.71 mmol/l

0.2 % at 40 ppm


0.2 % at 1.00 mmol/l

ACCURACY

1 % at 40 ppm


1 % at 1.74 mmol/l

1 % bei 40 ppm


1 % at 1.03 mmol/l

1 % at 40 ppm


1 % at 5.71 mmol/l

1 % at 40 ppm


1 % at 1.00 mmol/l

CALIBRATION

  • Linear with 2 standards
  • Nonlinear with 6-8 standards,
  • cubic approximation

DRIFT

1% in 60 min

REFERENCE

Lithium guide line with 35 mg/l or 5 mmol/l

SAMPLE VOLUME

2,5 ml

COMBUSTION GAS

Propane, Butane or LPG: Recommended for alkaline elements


Acetylene for flame photometry: Recommended for alkaline earth elements

DISPLAY

TFT-Display with integrated 8,4“TFT touchscreen, 800x600 Pixel

INTERFACES

  • 2 x USB
  • 1 x Ethernet
  • 1 x RS-232 for printer
  • Upgrade 4-20 mA analog, passive current interface

EQUIPMENT

Base unit + External valve box with control unit (Online Measurement)

ACCESSORIES

  • 24-character plain paper printer
  • Pressure reducing station with filter unit
  • Various standards and system solutions
  • Spare parts and accessories directly from the manufacturer

UPGRADEABILITY

Retrofit kits to other models available

Plus d’information
MEASURING RANGE Na 0.01-4500 ppm
Plage de mesure K 0.01-4500 ppm
MEASURING RANGE Li 0.01-4500 ppm
MEASURING RANGE Ca 0.50-4500 ppm
LIMITES DE DÉTECTION Na 0.01 ppm
LIMITES DE DÉTECTION K 0.01 ppm
LIMITES DE DÉTECTION Li 0.01 ppm
LIMITES DE DÉTECTION Ca 0.03 ppm
Précision du sodium. 0.2 % at 40 ppm
Précision du potassium. 0.2 % at 40 ppm
Précision du lithium. 0.2 % at 40 ppm
Précision du calcium. 0,2 % à 40 ppm.
PRÉCISION Na 1 % à 40 ppm
PRÉCISION K 1 % à 40 ppm
PRÉCISION Li 1 % à 40 ppm
PRÉCISION Ca 1 % à 40 ppm
Calibration Linéaire avec 2 étalons. Non linéaire avec 6-8 étalons, approximation cubique.
Volume d'échantillon 2,5 ml
Écran "Écran TFT intégré de 8,4 pouces avec écran tactile, 800x600 pixels
Interface 2 x clé USB 1 x Ethernet 1 x RS-232 pour imprimante Mise à niveau interface analogique 4-20 mA, courant passif
Rédigez votre propre commentaire
Vous commentez :Flame Photometer for Process FP8500
flamephoto (1.11 MB)